REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1024-R
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Event Description
It was reported to philips that there is physical damage to capnometer port.No patient harm or injury has been reported.Further information will be send upon completion of the manufacturer's investigation.
 
Manufacturer Narrative
The investigation confirmed that capnometer port is physically broken and charging port intermittently not charging.The capnometer has been replaced to address the issue reported initially.The device was operational after repairs were completed.The unit was returned to the customer site.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.A review of the risk management file was performed, and the potential severity is s2 has been identified in the risk document.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Type of Device
MONITOR, PHYSIOLOGICAL, PATIENT (WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key18039294
MDR Text Key326970595
Report Number3003832357-2023-00732
Device Sequence Number1
Product Code MHX
UDI-Device Identifier05060472441027
UDI-Public05060472441027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201746
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-1024-R
Device Catalogue Number989706000051
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received10/31/2023
Supplement Dates Manufacturer Received10/13/2023
Supplement Dates FDA Received12/28/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-