The customer declined the option to have their glidescope core smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined; however, it is likely that the reported damage to the cable's hdmi pins caused or contributed to the image issues.The glidescope video laryngoscopes operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged." verathon followed up with the customer and restated the importance of checking the device before its use in a procedure.Review of complaint history for the reported glidescope core smart cable serial number (b)(6) did not identify any previous complaints reported to verathon.Trending analysis for the glidescope core smart cable does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized by verathon.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
|
A customer reported that during a patient procedure, using a glidescope core smart cable, the image would disappear intermittently.The customer reported damage to the hdmi pins of the the smart cable.The procedure was completed using a backup glidescope system which was made available in an unspecified amount of time.No delay in the procedure or harm to the patient was reported.
|