Model Number 15 |
Device Problem
Electrical Power Problem (2925)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Stryker evaluated the customer¿s device, and was able to duplicate and verify the reported issue.Stryker determined that the cause of the reported issue was due to the power pcb assembly.The power pcb assembly was replaced to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
|
|
Event Description
|
The customer contacted stryker to report their device would flash on and then, reboot.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
|
Manufacturer Narrative
|
Upon return to the customer the problem reoccurred and further investigation found the root cause to be the system pcba.The system pcba was found to have an issue with crystal y1 intermittently not having an output.The system pcba was replaced to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
|
Event Description
|
The customer contacted stryker to report their device would flash on and then, reboot.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
|