MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problems Insufficient Cooling (1130); Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Manufacturer Narrative
H10: in order to solve the issue, patient bridge has been ordered and will be replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device took too long to cool down on the patient side.The issue occurred when the unit was in service.There was no patient injury.
 
Manufacturer Narrative
Quotation for a replacement patient pump 1 and distribution board was sent to customer and not accepted.Therefore, no service activity has been performed on the device.Complaints database review pointed out that no further similar issues have been submitted for this unit since its installation in 2020.Based on the above facts and on similar complaints investigations, it cannot be ruled out that the root cause of the issue is a corroded/stuck pump motor most likely due to environmental conditions, such as water infiltration, humidity, condensation or high temperatures, which required a power overload to work, that could have led to an overcurrent that caused electrical damage to the distribution board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18054259
MDR Text Key327133945
Report Number9611109-2023-00578
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public(01)04033817900528(11)201006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received11/02/2023
Supplement Dates Manufacturer Received02/16/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/06/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-