ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Device Problem Computer Software Problem (1112)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event Description
The customer reported that cbct shifts are misaligned when imported to mosaiq.
 
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported that the cone beam computed tomography (cbct) shifts are misaligned when imported to mosaiq.It was ascertained that when the cbct was sent back to mosaiq for recording, the offsets were no longer accurate.The recommended workflow allows the customer to verify the shifts seen on truebeam and compare with the image list in mosaiq to verify accuracy.This is a detectable post treatment recording issue not affecting or leading to patient mistreatment.Elekta were unable to reproduce the issue and are therefore unable to determine a root cause for this issue.This is believed to be a one-time occurrence and the customer has not reported any reoccurrence of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key18054891
MDR Text Key327140987
Report Number3015232217-2023-00068
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002367
UDI-Public(01)00858164002367(10)2.83.040
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received11/02/2023
Supplement Dates Manufacturer Received10/04/2023
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse