Model Number FV414T
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Hydrocephalus (3272)
Event Type  Injury  
Manufacturer Narrative
Visual inspection: visual inspection did not reveal any significant damage, but did detect deformation of the housing surface.Measurement of the plane parallelism showed a deformation with a value of -0.073 mm [outside the tolerance of 0 ± 0.02 mm].Permeability test: a permeability test has shown that all components are permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the both valves are not operating within the acceptable tolerance in either position.An accelerated outflow of both valves could be determined.Adjustment test: the progav was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in both valves.To make the proteins / deposits in the shunt system more visible, they were colored using a staining solution.Results: based on our investigation results, we can determine an accelerated discharge at the valves, in their respective relevant position.The deposits visible in the valves may have led to the change in flow rate.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event Description
It was reported that a progav shunt system (#fv414t) was implanted during a procedure performed on unknown according to the complainant, the shunt showed a over-drainage and adjustment difficulties.The patient underwent a revision procedure performed on unknown.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 1 years.Weight: 11 kilograms.Height: 105 centimeters.Gender: female.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV WITH SHUNTASSISTANT 25
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18055421
MDR Text Key327169957
Report Number3004721439-2023-00321
Device Sequence Number1
Product Code JXG
UDI-Device Identifier04041906132118
UDI-Public4041906132118
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K103003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFV414T
Device Catalogue NumberFV414T
Device Lot Number20059542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2023
Initial Date FDA Received11/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age12 MO
Patient SexFemale
Patient Weight11 KG
-
-