An investigation has been initiated.We are currently waiting for additional information regarding the incident and for the return of the catheter for evaluation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Received one intact 13f x 15cm duo-split catheter for evaluation.A visual evaluation denoted no obvious defects.The extension tubing is slightly discolored.There is no damage or leakage from the red luer.The blue luer has a visible crack 1.0cm in length and appears to be next to the luer knit line.When flushing the device, a small stream leaks from the crack.No leakage occurs when the device is not under pressure.Under magnification, there appears to be damage to the luer wings that may have been caused by the use of an instrument on the luer.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.We are unable to determine the cause or factors that may have contributed to this event.The female luers are manufactured to meet the iso standard.A common cause of cracked luers is over-tightening of connections.The ergonomically shaped luer provides users with a patent locking design.This assures obtaining a secure mating of the catheter without the need to torque down.
|