Investigation: visual inspection: during the investigation, some residue tissues on the device was determined.No significant deformations or damage of the valve were detected during the visual inspection, the measurement of the plane parallelism confirmed it with a value of -0.017 mm [within the tolerance of 0 ± 0.02 mm].Permeability test: a permeability test has shown that all components are permeable.During the permeability test bloody liquid were flushed out.Computer control test: to investigate the drainage, the opening pressure is measured using a miethke computer control testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav is operating not within the accepted tolerance in the horizontal position.The shuntassistant is operating within the specified tolerances in the vertical position.An accelerated outflow of progav could be determined.Adjustment test: the progav was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection : after dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we can determine an accelerated outflow and adjustment difficulties on the progav valve.The determined deposits can be named as the cause for the functional deviations.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|