Model Number 866389 |
Device Problem
Alarm Not Visible (1022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported that the system has a lack of red alarms.The device was in use on a patient.There was no report of patient or user harm.
|
|
Manufacturer Narrative
|
Visual inspection found the device did alarm yellow.Results of functional testing indicate wide beats learned as normal.The device was operational after rse was able to show tech why the monitor did not alarm properly.Monitor keeps losing leads and relearning beats.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|