Model Number P6881 |
Device Problems
Material Frayed (1262); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The baxter technician found the power cord needed to be replaced.Per the hillrom service manual inspect power cord for damage, twisting and replace if necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed on (b)(6) 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
|
|
Event Description
|
Baxter received a report from a baxter technician stating the bed had power cord damage with frayed exposed copper wires.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
|