• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE X EP SYSTEM SURFACE ELECTRODE KIT (5PK US); COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE X EP SYSTEM SURFACE ELECTRODE KIT (5PK US); COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-SEK-5-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Skin Tears (2516)
Event Date 10/16/2023
Event Type  Injury  
Event Description
During an atrial fibrillation procedure, the patches were very sticky and the patient's skin minorly teared under the left armpit upon removal.The area had been prepped for patch placement including shaving hairy areas.No treatment was necessary for the skin tear and there were no alleged performance issues with the patches.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported skin tear could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE X EP SYSTEM SURFACE ELECTRODE KIT (5PK US)
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18057718
MDR Text Key327195064
Report Number3008452825-2023-00506
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067032379
UDI-Public05415067032379
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K231415
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberENSITE-SEK-5-01
Device Lot Number9086950
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received11/02/2023
Supplement Dates Manufacturer Received11/20/2023
Supplement Dates FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-