MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH SCP WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH SCP WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-40-00
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a report that a s5 console displayed error erc timout (e117) in the scp with all display freeze during a procedure.There was no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the scp with tubing clamp.The incident occurred in france.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H10: e117 signals a timeout of the erc and can be caused by: contact errors on plug-in connectors, defective main connector cable, trip of the circuit breaker of the tubing clamp, power supply issue that prevent the connection with scp system.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.Can internal and external communication boards were replaced without solving the issue.Therefore, the whole faulty e/p (electronics and power) pack was replaced with a loan one and the issue was then fixed.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2014.Based on technical intervention, most likely root cause of the event could be traced back to a failure of an internal component of the e/p pack, likely due to the wearing of the unit that prevented the erc to be properly power supplied and therefore losing connection with the scp system.Failure of electronic components can be related to multiple and not deterministic factors such as exposure to heat, dust and moisture, accidental impacts (drops and falls), and power overloads/surges but also to variability of micro sub-components.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCP WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18064749
MDR Text Key327273122
Report Number9611109-2023-00588
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received11/03/2023
Supplement Dates Manufacturer Received10/13/2023
Supplement Dates FDA Received01/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-