Model Number 1944/52
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure it wad noted that the right ventricular lead exhibited failure to sense.The lead was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of failure to sense was not confirmed.A complete lead was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection and x-ray examination of the lead did not find any anomalies.All tines were intact and undamaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18067241
MDR Text Key327311404
Report Number2017865-2023-50953
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734502320
UDI-Public05414734502320
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1944/52
Device Lot NumberP000171770
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2023
Initial Date FDA Received11/03/2023
Supplement Dates Manufacturer Received01/09/2024
Supplement Dates FDA Received01/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-