The device was received and evaluated.There were no external visual defects noted.A functional test was performed using a stellaris system.The handpiece data displayed tune count 19, on-time 1455892 and average power 0.The handpiece tuned, primed, and functioned as intended.The device history record was received and there were no nonconformities or anomalies related to this complaint.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.The root cause of the reported event could not be determined.
|