• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedure, the aquabeam robotic system generated an "e22 - motorpack error" message, which could not be cleared despite multiple troubleshooting steps taken in an effort to resolve the issue.A second aquabeam handpiece was used, which successfully resolved the issue, and the aquablation procedure was completed successfully.The reported event caused a surgical delay of over 20 minutes.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The aquabeam handpiece was returned for investigation.Visual inspection of the returned device revealed no physical damage or anomalies.Functional testing of the device could not confirm the reported event.The handpiece was deconstructed and viewed under magnification.No signs of fluid ingress were observed on the sensor board and the encoder wheel.The root cause of the reported event is undeterminable as the handpiece functioned as intended.The aquabeam robotic system's treatment logs file was reviewed which confirmed the reported event.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c03114 was conducted, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were reworked to address the nonconformance.Upon re-inspection, the lot met all required specifications and was then deemed acceptable to be released for distribution per device specifications.The current user manual um0101-00 rev.F, aquabeam robotic system user manual, us, english was reviewed.Table 5 system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.E23 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece e50 - console error release foot pedal and click x.If error persists, turn off and turn on console and cpu submission of this report does not constitue an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech dr
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech dr
san jose, CA 95134
6502327291
MDR Report Key18069936
MDR Text Key328143951
Report Number3012977056-2023-00184
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20230320D
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/11/2023
Initial Date FDA Received11/03/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received12/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-