• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM5500
Device Problems Failure to Interrogate (1332); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the clinic for a scheduled follow up.Upon interrogation of the implantable cardiac monitor (icm), the bluetooth low energy (ble) was unstable and eventually the ble telemetry was lost with a prompt of an error message.The physician attempted several times to interrogate the icm but was unsuccessful.The patient was in stable condition.
 
Manufacturer Narrative
Further information was requested but not received.
 
Manufacturer Narrative
The reported event of an inability to interrogate the device resulting in a programmer crash was confirmed.Based on the information provided and reviewed by abbott engineering, it was determined that the programmer crash was due to a programmer code not setting the segmentation related session attribute to 0 before requesting a bulk attribute in a new session.The device is performing per design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSERT-IQ¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18071083
MDR Text Key327368187
Report Number2017865-2023-51688
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067047809
UDI-Public05415067047809
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K230286
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM5500
Device Lot NumberS000091360
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/17/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-