• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE PHACO TIP; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE PHACO TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number ASKU
Device Problem Obstruction of Flow (2423)
Patient Problem Eye Burn (2523)
Event Date 10/04/2023
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during a cataract surgery an ophthalmic handpiece exhibited thermal issue due to occlusion.Tip heated causing thermal burn at incision site.Suture was used to close the wound.Patient condition was resolved.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of thermal issue, nylon sutures used to close wound, occlusion during surgery; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHACO TIP
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18071709
MDR Text Key327354797
Report Number2523835-2023-00616
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/11/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received01/30/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age74 YR
Patient SexFemale
-
-