• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOS 3.5MM HEX DRIVE; INSTRUMENT, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOS 3.5MM HEX DRIVE; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).D10: 31-301858 item name arcos taper disasmbly 70mm lot # zb180502.Multiple mdr reports were filed for this event, please see associated reports 0001825034 - 2023 - 02566.The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported that the disassembly bolt worked as intended during surgery.When the procedure was completed, the disassembly bolt from the proximal body had cross threaded.The proximal body-tip of 3.5 hex driver also broke trying to provide counter torque to release disassembly bolt from proximal body.The patient was no affected in any way.There is no additional information available at the time of this report.
 
Event Description
There is no update to the prior event description provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the tip of the hex driver to be fractured.The tip was not returned.Rings of wear have been scuffed around the shaft.Discoloration spots are present on the grip.The connector is lightly nicked and scratched.The features of the fracture surface visually align with zrm in which a bending overload fracture failure mode was identified.Complaint confirmed based on the evaluation of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 3.5MM HEX DRIVE
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18072353
MDR Text Key327377866
Report Number0001825034-2023-02565
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00887868477898
UDI-Public(01)00887868477898(11)150901(10)ZB150701
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301852
Device Lot NumberZB150701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
PLEASE SEE H10.
Patient SexMale
-
-