Model Number N/A |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/09/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).D10: 31-301858 item name arcos taper disasmbly 70mm lot # zb180502.Multiple mdr reports were filed for this event, please see associated reports 0001825034 - 2023 - 02566.The product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that the disassembly bolt worked as intended during surgery.When the procedure was completed, the disassembly bolt from the proximal body had cross threaded.The proximal body-tip of 3.5 hex driver also broke trying to provide counter torque to release disassembly bolt from proximal body.The patient was no affected in any way.There is no additional information available at the time of this report.
|
|
Event Description
|
There is no update to the prior event description provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the tip of the hex driver to be fractured.The tip was not returned.Rings of wear have been scuffed around the shaft.Discoloration spots are present on the grip.The connector is lightly nicked and scratched.The features of the fracture surface visually align with zrm in which a bending overload fracture failure mode was identified.Complaint confirmed based on the evaluation of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|