MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-M
Device Problems Material Deformation (2976); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Event Description
A hawkone device was used for treatment of a fibrous plaque lesion in the mid right superficial femoral artery.The artery was 6mm in diameter with mild tortuosity and moderate calcification.A 6fr non medtronic sheath was used and a 0.14 non medtronic guidewire.The vessel was not pre-dilated.The ifu was followed.It was reported after one pass, turned off to pack and nosecone looked 3/4 full, turned on again for another pass, packed again and device was then only 1/4 full.Took device out of body to clean and looked at nosecone and plaque was coming out of side of nosecone, the plaque looked to be bulging out of side.The device had made enough lumen gain for a dcb to complete the procedure.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis: the device was returned with the distal flush tool (dft) positioned over the nosecone.The dft was retracted, and a visual inspection found a bulge on the distal end of the nosecone and biologics in the nosecone.The tecothane appears to be split.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18072992
MDR Text Key327397171
Report Number9612164-2023-05038
Device Sequence Number1
Product Code MCW
UDI-Device Identifier00643169968332
UDI-Public00643169968332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH1-M
Device Lot Number0011800328
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received11/30/2023
Supplement Dates FDA Received12/05/2023
Date Device Manufactured05/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
-
-