Catalog Number CDS0705-XTW
Device Problems Difficult to Insert (1316); Retraction Problem (1536); Unintended Movement (3026); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All available information was investigated and without the device to analyze, a cause for the reported unintended movement of the clip, difficult to insert and retract the clip into/from the clip introducer (ci) resulting in torn ci could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4 and restricted posterior leaflet.A mitraclip xtw was advanced to the left atrium and after establishing the straddle position and while adding m knob at 90 degrees, the clip exhibited miskey behavior where it moved in the cranial position and was retracted.It was noted that that there had been strong resistance while fitting the clip into the introducer and became stuck between the clip arm and the shaft.The clip was advanced and canceled the interference, and the clip was able to be pulled back.The clip fit in without any resistance.The device was removed from the patient.After taking the device out of the body, the tip of the clip introducer was examined and a hole at the tip of the clip introducer was observed.The procedure was completed with two clips implanted.The mr was reduced to trace.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18074925
MDR Text Key327424029
Report Number2135147-2023-04880
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/03/2024
Device Catalogue NumberCDS0705-XTW
Device Lot Number30405R1042
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received11/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER.
-
-