Catalog Number CDS0705-XTW |
Device Problems
Difficult to Insert (1316); Retraction Problem (1536); Unintended Movement (3026); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All available information was investigated and without the device to analyze, a cause for the reported unintended movement of the clip, difficult to insert and retract the clip into/from the clip introducer (ci) resulting in torn ci could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Event Description
|
It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4 and restricted posterior leaflet.A mitraclip xtw was advanced to the left atrium and after establishing the straddle position and while adding m knob at 90 degrees, the clip exhibited miskey behavior where it moved in the cranial position and was retracted.It was noted that that there had been strong resistance while fitting the clip into the introducer and became stuck between the clip arm and the shaft.The clip was advanced and canceled the interference, and the clip was able to be pulled back.The clip fit in without any resistance.The device was removed from the patient.After taking the device out of the body, the tip of the clip introducer was examined and a hole at the tip of the clip introducer was observed.The procedure was completed with two clips implanted.The mr was reduced to trace.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Search Alerts/Recalls
|
|