• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CORTSCR PLUSDRIVE Ø2 SELF-DRILL L6 TAN; SCREW, FIXATION, INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CORTSCR PLUSDRIVE Ø2 SELF-DRILL L6 TAN; SCREW, FIXATION, INTRAOSSEOUS Back to Search Results
Catalog Number 401.063.99
Device Problem Nonstandard Device (1420)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2023
Event Type  malfunction  
Event Description
Device report from synthes reports an event in china as follows: it was reported that during incoming inspection by distributor, it was found that some screws have partially missing threads.There were no adverse consequences to patient.No additional information could be provided.This report is for one (1) cortscr plusdrive ø2 self-drill l6 tan this is report 2 of 10 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h4, h6 the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device visual analysis of the photo revealed that the cortscr plusdrive ø2 self-drill l6 tan threads are defective.Event though only one photo was received, this is representative of the confirmed issue.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for cortscr plusdrive ø2 self-drill l6 tan.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H4, h6 manufacturing location: monument manufacturing date: 04-feb-2022 part number: 401.063.99, 2.0mm ti cortex screw slf-drlg w/plusdrive ¿ recess 6mm lot number: 607p102 (non-sterile) production order traveler met all inspection acceptance criteria.Inspection sheet, inspect dimensional/final inspection, ns034688 rev ab met all inspection acceptance criteria.Packaging label log (pll) lmd rev ad was reviewed and determined to be conforming.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿missing threads¿ does not indicate breakage.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTSCR PLUSDRIVE Ø2 SELF-DRILL L6 TAN
Type of Device
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18075445
MDR Text Key327428813
Report Number8030965-2023-13961
Device Sequence Number1
Product Code DZL
UDI-Device Identifier07611819865228
UDI-Public(01)07611819865228
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K170818
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number401.063.99
Device Lot Number607P102
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received10/19/2023
Supplement Dates FDA Received11/17/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/04/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-