Reportable based on device analysis completed on (b)(6) 2023.It was reported that the system alerted an error, and a tube leak occurred.An angiojet ultra pe was selected for used for thrombectomy procedure.During the procedure, it was observed that there was an infusion error during the initiation of the first three pulses.Additionally, there was a catheter connection leakage, prompting a message to change the device.The procedure was completed with a different device.There were no patient complications reported.However, returned device analysis revealed a catheter shaft kink/fracture.
|
Device evaluated by mfr: the product was returned to boston scientific for analysis.Returned product consisted of an angiojet pe catheter.The photo was reviewed which showed the alleged leak and was confirmed during analysis.The catheter shaft showed a kink/fracture at 73 cm from the tip.Functional testing was completed per the ifu device preparation.During functional testing no errors or priming issues were noticed, the device ran as designed.During functional testing there was a leak at the male luer connector.During functional testing no errors or priming issues were noticed, the device ran as designed.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was not confirmed for set up issues and alarms.The device was confirmed the male luer connector leak and a shaft kink/fracture.
|