ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Incorrect Interpretation of Signal (1543)
Patient Problem Insufficient Information (4580)
Event Date 10/24/2023
Event Type  Death  
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Manufacturer Narrative
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5.Evaluation: the device was returned to a zoll medical corporation for evaluation.The customer's report was verified during the internal inspection where the primary side shield was observed to be loose.The loose shield likely contributed to noise immunity and susceptability.The analog board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable due to a noisy ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18077273
MDR Text Key327440598
Report Number1220908-2023-04396
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/27/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received01/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-