Model Number R SERIES |
Device Problem
Incorrect Interpretation of Signal (1543)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/24/2023 |
Event Type
Death
|
Event Description
|
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Manufacturer Narrative
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section b5.Evaluation: the device was returned to a zoll medical corporation for evaluation.The customer's report was verified during the internal inspection where the primary side shield was observed to be loose.The loose shield likely contributed to noise immunity and susceptability.The analog board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|
Event Description
|
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable due to a noisy ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
|
Search Alerts/Recalls
|