Catalog Number IAV09008008P |
Device Problems
Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/23/2023 |
Event Type
Injury
|
Event Description
|
Physician attempted to use an inpact av access drug coated balloon with a 7fr non-medtronic sheath during treatment of a fibrous lesion in the patients mid brachial cephalic artery.No vessel tortuosity and no vessel calcification are reported.Embolic protection was not used.A non-medtronic inflation device was used for balloon inflation.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.There were no issues noted when removing the device from the hoop/tray.Ifu was followed and the device was prepped without issue.The device did not pass through a previously-deployed stent.No resistance was encountered when advancing the device.Excessive force was not used.It is reported the balloon broke apart and the tip separated during balloon inflation at rbp.Pieces of the balloon had to be snared from the artery and the non-medtronic sheath had to be removed to pull the broken balloon from the patient, and access was lost.All fragments of the balloon were retrieved.No patient injury reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: the procedure was aborted after access was lost.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis the device returned with a balloon and inner member detachment, a detached section of the inner member and a detached section of the balloon.The balloon was split 0.6cm to 2.3cm distal to the proximal balloon bond, and was detached 5.3cm distal to the proximal balloon bond.The balloon detachment site was smooth but uneven.The small, detached, 1.5cm section of the balloon was held tightly together with solidified blood.There was an inner member detachment 9.3cm distal to the proximal balloon bond and the detachment site was jagged and uneven.The detached, 9.0cm section of the inner was stretched along its entire length and the detachment sites at either end were uneven and jagged.A tactile test detected a kink to the catheter at 15.2cm distal to the strain relief.The distal tip was not returned.The balloon could not be inflated due to the detachment.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|