This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, the device was not returned to olympus and the event, ¿distal end of the device came off¿, was not confirmed.The root cause could not be identified.However, the subject device was repaired by third party agency, and it was probable that repair against instructions for use (ifu) was performed.Ifu prohibits repair by third party agency other than olympus.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 13 years since the subject device was manufactured.The instruction manual identifies the following related verbiage which could have prevented the phenomenon: ¿instruction_ important information ¿ please read before use_ prohibition of improper repair and modification: this instrument does not contain any user-serviceable parts.Do not disassemble, modify, or attempt to repair it; patient or operator injury and/or equipment damage can result.Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than olympus¿ own authorized service personnel is excluded from olympus¿ limited warranty and is not warranted by olympus in any manner.¿ olympus will continue to monitor field performance for this device.
|