Catalog Number 19AGFN-756
Device Problems Fracture (1260); Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Manufacturer Narrative
An event of leaflets stuck close and fractured leaflet was reported.The device was returned to abbott for investigation and one leaflet was dislodged and fractured while the other leaflet was properly inserted into the orifice.The investigation found the orifice was chipped.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.While the root cause of the leaflet dislodgement could not be conclusively determined, there was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflets or orifice damage.The damage may have been caused by some external force applied to the valve which overstressed the carbon material.Please note, per the regent valve instructions for use, "valve rotation: using the valve holder/rotator and the flexible valve holder handle model 905-hh, or the rigid valve holder handle model 905-rhh, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation".
 
Event Description
It was reported that on (b)(6) 2023, a 19mm regent aortic mhv,flex c uff, was selected for an implant.During implantation the valve was cor-knotted in and then rotated.When the holder was removed the leaflets were stuck closed.They did everything possible to open the leaflets.When they got one leaflet to open the other leaflet broke into three pieces in the left ventricle and was recovered from the patient.The leaflets and valve were removed and a non abbott valve was implanted.The patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18088126
MDR Text Key328026479
Report Number2135147-2023-04905
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number19AGFN-756
Device Lot NumberC00011803
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received11/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-