IMRIS - DEERFIELD IMAGING, INC. IMRIS ORT300 OPERATING ROOM TABLE Back to Search Results
Model Number ORT300
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Manufacturer Narrative
The imris customer service engineer investigated the issue on-site and observed a small hydraulic fluid leak from a pressurized input hose associated with the back lift cylinder.The cse measured the hydraulic fluid level in the operating room table and confirmed it was within specification.The table was functionally tested and no functional issues were present or previously reported by the customer.Replacement of the hydraulic hose is in process; a follow-up report will be submitted upon completion of the investigation.
 
Event Description
A biomedical technician at the customer site reported observing hydraulic fluid at the base of the ort300 operating room table.There was no patient or procedural involvement at the time of the observation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT300 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
1230 chaska creek way
suite 100
chaska MN 55318
Manufacturer Contact
paul campbell
1230 chaska creek way
suite 100
chaska, MN 55318
7632036344
MDR Report Key18088480
MDR Text Key328498738
Report Number3010326005-2023-00012
Device Sequence Number1
Product Code KXJ
UDI-Device Identifier00857534006356
UDI-Public(01)00857534006356
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberORT300
Device Catalogue Number114093-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/09/2023
Initial Date FDA Received11/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-