Model Number CRPLUS
Device Problem Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2023
Event Type  malfunction  
Event Description
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device had logged an event code in the memory which is related to a potential issue of the device to deliver energy.There was no patient involvement reported with the event.
 
Manufacturer Narrative
Stryker evaluated the customer's device.During evaluation the device shocked as expected before and after clearing error codes.Further investigation could not isolate a root cause for the reported issue.This device was scrapped and the customer received a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® CR PLUS DEFIBRILLATOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
dara friedman
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key18089944
MDR Text Key327591298
Report Number0003015876-2023-02047
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873823644
UDI-Public00883873823644
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99427-000129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2023
Initial Date Manufacturer Received 10/09/2023
Initial Date FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-