MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. FEMORAL STEM 12/14 NECK TAPER EXTRA EXT. OFFSET; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. FEMORAL STEM 12/14 NECK TAPER EXTRA EXT. OFFSET; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fall (1848); Bone Fracture(s) (1870)
Event Date 10/13/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10: 00877503601 item name bioloxâ® delta, ceramic femoral head, s, 㸠36/-3.5, taper 12/14 lot # unk.G2: foreign: australia.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted h3 other text : device location is unknown.
 
Event Description
It was reported that the patient underwent a revision procedure 4 years post implantation due to periprosthetic fracture that was secondary to a fall.There is no additional information available at the time of this report.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Proposed component code: mechanical (g04): stem.Visual examination of the provided picture identified the explanted head and stem, covered in bio-debris.No further evaluation can be made from the provided picture.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: left tha with periprosthetic fracture of the medial aspect of the proximal femoral diaphysis.It was reported the patient suffered a periprosthetic fracture due to a fall.However, as the reason for the fall is unknown, a definitive root cause cannot be determined.This complaint was confirmed based on the provided x-rays.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER EXTRA EXT. OFFSET
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18092409
MDR Text Key327631807
Report Number0002648920-2023-00265
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K191735
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/31/2024
Device Model NumberN/A
Device Catalogue Number00811400430
Device Lot Number62716029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received11/08/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
PLEASE SEE H10
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age68 YR
Patient SexMale
Patient Weight85 KG
-
-