• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT CONFIRM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT CONFIRM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM2102
Device Problems Failure to Interrogate (1332); Incorrect, Inadequate or Imprecise Result or Readings (1535); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Event Description
During an in-clinic follow-up, the device was unable to be interrogated and delivered an error message.In addition, previous episodes of undersensing were also observed on the device.Abbott technical support was contacted and suspected a corrupt egm was the cause of the event.No intervention was performed at this time.The patient was stable and will continue to be monitored.
 
Event Description
Additional information received indicated the patient was later seen in clinic, and the device was reprogrammed to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18093026
MDR Text Key327639476
Report Number2017865-2023-51905
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05414734502573
UDI-Public05414734502573
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K133481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model NumberDM2102
Device Lot NumberS000084208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/08/2023
Supplement Dates Manufacturer Received11/14/2023
Supplement Dates FDA Received11/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-