There was an allegation of questionable free psa elecsys results for one patient from the cobas e411 rack analyzer that were higher than the total psa result for the sample.Sample 1 initial free psa result was 0.058 ng/ml.After replacing the reagent pack, calibration and qc, the repeat result was 0.176 ng/ml.Sample 2 initial free psa result was 0.153 ng/ml.After replacing the reagent pack, calibration and qc, the repeat result was 0.15 ng/ml.The sample was sent to another laboratory and tested using clia methodology.The free psa result was 0.06 ng/ml.The result from the other laboratory was believed to be correct.The questionable results were not reported outside of the laboratory.
|