MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN POLY EXTRACTOR; EXTRACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN POLY EXTRACTOR; EXTRACTION INSTRUMENTS Back to Search Results
Catalog Number 221750001
Device Problem Material Deformation (2976)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the instrument is bent and unusable.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : it was reported that the instrument is bent and unusable.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the extractor head was found bent from the pinn poly extractor.The complaint condition can be confirmed.The device exhibits damage consistent with repeated use over 18 years.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the pinn poly extractor would contribute to the complained device issue.Based on the investigation findings, potential cause can be traced to end of device life and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code provided is not a finished goods lot number.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: an analysis of the product could not be performed, since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected and distributed to approved specifications.Additional complaint information, monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN POLY EXTRACTOR
Type of Device
EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18097855
MDR Text Key327755434
Report Number1818910-2023-22900
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295098928
UDI-Public10603295098928
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750001
Device Lot NumberJ0305
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2023
Initial Date FDA Received11/08/2023
Supplement Dates Manufacturer Received11/30/2023
12/28/2023
Supplement Dates FDA Received12/04/2023
12/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-