• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KNEE SCORPION; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KNEE SCORPION; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number KNEE SCORPION
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Insufficient Information (4580)
Event Date 10/16/2023
Event Type  malfunction  
Event Description
On 17th october 2023 it was reported by a sales rep via email that an ar-12990, (knee scorpion¿) was unable to completely close and upon closer inspection of the device the spring that allows the trigger to recoil had become dislodged and blocked the ability to close the jaws.This was discovered during a meniscal root repair on (b)(6) 2023.The surgeon successfully completed the procedure by using the other consigned knee scorpion to complete the procedure.
 
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE SCORPION
Type of Device
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18100478
MDR Text Key327721040
Report Number1220246-2023-08620
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00888867196322
UDI-Public00888867196322
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKNEE SCORPION
Device Catalogue NumberAR-12990
Device Lot Number10253200
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received11/09/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-