Model Number 15 |
Device Problem
Electrical Power Problem (2925)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A stryker service representative performed an initial evaluation of the customer¿s device and was able to verify the reported issue.After replacing the system pcb, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
|
|
Event Description
|
The customer contacted stryker to report that their device was not booting up when powered on.This issue has the potential to either delay, or prevent, defibrillation from being delivered.There was no patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Stryker further evaluated the replaced part at the product assessment center (pac) and the cause of the reported issue was determined to be due to short circuit on the system pcb assembly.
|
|
Event Description
|
The customer contacted stryker to report that their device was not booting up when powered on.This issue has the potential to either delay, or prevent, defibrillation from being delivered.There was no patient use associated with the reported event.
|
|