MEDTRONIC IRELAND VALIANT CAPTIVIA - CW; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
Model Number VAMC3834C150TJ |
Device Problems
Difficult to Insert (1316); Device Damaged by Another Device (2915)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2023 |
Event Type
malfunction
|
Event Description
|
Valiant captivia stent graft (vamc3834c150tj) was intended to be implanted during a two-part endovascular treatment of an 53mm thoracic aortic aneurysm.The patient had undergone a y graft replacement approximately 20 years previously , where the blood vessel diameter measured 8mm. it was determined that there would be no issues with passing through the delivery system so the procedure went ahead as usual.It was reported during the index procedure, when the peripheral device vamc3834c150tj was inserted in the right main, it did not advance and became stuck in the peripheral anastomotic region of the y graft.The graft was withdrawn and was inserted in the left side which was newly cut down, the device advanced further than the right but stopped at the leg of the y graft.Multiple attempts were made to advance the device, but it was ultimately decided to withdraw the delivery system.This decision was based on concerns that the device might become irretrievable if advanced further in the first run or that the second run might not be successful. it was noted that non-mdt devices , which are lower profile devices than valiant captiva, were not available as back ups at the time.Per the physician the cause of the delivery issue was related to the fact that 20 years that passed since the placement of the y graft which made this procedure more difficult.No additional clinical sequelae were provided, and the patient will be monitored.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device evaluation summary: the device returned with the external slider in the home position.A kink was visible on the graft cover 8.7cm from the taper tip.There was no further deformation evident to the device.The reported insertion difficulties were confirmed through device analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|