U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD2411-36Q
Device Problem Failure to Convert Rhythm (1540)
Patient Problem Arrhythmia (1721)
Event Date 10/20/2023
Event Type  Injury  
Event Description
Related manufacturer reference number: (b)(4).It was reported, that the patient presented in clinic, due to an arrhythmia.Device interrogation revealed, that it was unable to convert the rhythm needing external defibrillation on the implantable cardioverter defibrillator (icd).An x-ray confirmed dislodgement on the atrial (ra) lead.No changes or intervention were reported.The patient was in stable condition.
 
Event Description
New information noted that both the implantable cardioverter defibrillator (icd) and the atrial (ra) lead was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18108077
MDR Text Key327865502
Report Number2017865-2023-52058
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507615
UDI-Public05414734507615
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2016
Device Model NumberCD2411-36Q
Device Lot Number4485427
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/09/2023
Supplement Dates Manufacturer Received11/07/2023
04/30/2024
Supplement Dates FDA Received11/29/2023
04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7121Q/65
Patient Outcome(s) Other; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-