• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.A getinge technician will investigate the hl 20.A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the hl 20 pump displayed the error message "beltslip" and after some time also the error message ¿error head¿.The event occurred during start up.No harm to any person was reported.Complaint id: (b)(4).
 
Manufacturer Narrative
It was reported that the hl 20 pump displayed the error message "beltslip" and after some time also the error message ¿error head¿.The event occurred during start up.No harm to any person was reported.A getinge field service technician (fst) was sent for investigation and repair on 2023-11-19.The optical tacho board was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The reported failures "error head" was already investigated in a similar complaint by the getinge life cycle engineering (lce) on 2019-08-22.The most probable root cause was determined as a broken wiring of the signal line j1-3 on the optical tacho board.According to lce assessment from 2024-01-11 the error "beltslip" can also be caused by the same broken wiring.Therefore the root cause is most probably the same.The review of the non-conformities has been performed on 2023-11-09 for the period of 2011-09-04 to 2023-11-03.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2011-09-04.Based on the results the reported failure "error message "beltslip" and ¿error head¿" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18109067
MDR Text Key328484283
Report Number8010762-2023-00561
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701024771
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2023
Initial Date FDA Received11/10/2023
Supplement Dates Manufacturer Received01/11/2024
Supplement Dates FDA Received01/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-