BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-2202-45
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Undesired Nerve Stimulation (1980); Burning Sensation (2146)
Event Date 08/19/2021
Event Type  Injury  
Event Description
It was reported that the deep brain stimulation (dbs) patient experienced side effects in the eye when stimulation was on causing burning sensations.The physician attempted to try two different programs to address the eye burning.One where the patient had more severe tremor but better effect on the eye burning, and another where the tremor was better but the eye burning would occur.Mri imaging was performed and revealed the leads were sub-optimally placed.The patient underwent a revision procedure where the leads were replaced.The patient was doing well post-operatively.The explanted leads were retained by the hospital and were not returned to bsc.
 
Manufacturer Narrative
Additional suspect medical device component involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(6), batch: 7081903.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE CARTESIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18112807
MDR Text Key327868900
Report Number3006630150-2023-06942
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729905288
UDI-Public08714729905288
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/08/2023
Device Model NumberDB-2202-45
Device Catalogue NumberDB-2202-45
Device Lot Number7081899
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2023
Initial Date FDA Received11/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexMale
-
-