MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INLET, MICROVOL,VENTED SPIKE,; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INLET, MICROVOL,VENTED SPIKE,; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938751
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Event Description
It was reported that the spike of a micro-volume inlet detached resulting in a leak.This was observed during compounding.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured from september 29, 2022 - october 03, 2022.H10: the device was received for evaluation.Unaided eye visual inspection was performed and the inlet spike was received detached from the inlet tubing.Functional testing was not possible as the sample was damaged.The reported condition was verified.The cause of the condition could not be determined; however, the most likely cause was due to inadequate or lack of solvent being applied to the outer diameter of tube when it was inserted to the spike in the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INLET, MICROVOL,VENTED SPIKE,
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18115715
MDR Text Key328387986
Report Number1416980-2023-05846
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412046341
UDI-Public(01)00085412046341
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K903159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938751
Device Lot Number60400405
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2023
Initial Date FDA Received11/10/2023
Supplement Dates Manufacturer Received12/14/2023
Supplement Dates FDA Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-