This report is based on information provided by a philips field service engineer and a bench engineer and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro indicating that the touchscreen was not working.Diagnostics determined the device would require return for repair and it was returned to a philips bench repair facility where the reported complaint was confirmed as the trizeps 7 board was dislodged from the sbc (single board computer).The sbc was replaced resolving the customer¿s complaint (rdt sbc kit (for 00-102x-r & 00-103x-r), part 453564865071).Testing and verification was completed satisfactorily (calibrated nbp, co2 and touch screen) and the device was returned to service at the customer site.Based on the information available and testing conducted, the cause of the reported problem was the trizeps 7 board becoming dislodged.The reported problem was confirmed.Based on the information available, no further action is necessary at this time.While no harm was alleged, recurrence of this failure mode during clinical use could cause or contribute to death or serious injury.Therefore, this failure mode meets the definition of a reportable malfunction.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|