Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  malfunction  
Event Description
It was reported that the brake failed.The patient presented with angina and required a rotational atherectomy procedure.During removal of a rotapro device while in dynaglide mode, the "burr vibrates" indicating a console speed issue.The advancer brake failed to lock on the guidewire and the guidewire couldn't be controlled forcing both the burr and wire out of the patient's body.The procedure was completed with the original device and no patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18119704
MDR Text Key327930150
Report Number2124215-2023-63586
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2023
Initial Date FDA Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-