LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that during a following up, when interrogating the generator error code 10 (burst watchdog timeout) was observed.It was noted that the autostim was off.Tablet data was also provided for review.The provider also added that the patient was not exposed to any high energy fields, did not undergo any recent medical procedures, and did not experienced any environmental changes that may impact the ipg.A burst watchdog timeout reset is a risk mitigation to ensure the therapy (on time) is not delivered for the longer duration and intended to cover unexpected scenarios.Potential causes of burst watchdog timeout for this device could be a memory error (which could not be detected as the failure signature is not recorded in the file) or a possible firmware bug.The generator data decoder confirms that the burst watchdog timeout was encountered when the device was interrogated.According to the data, therapy was reenabled following the timeout error and system diagnostics were within normal limits after event.No other relevant information has been received to date.
 
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Update was received that the generator was interrogated and a therapy disabled message was seen again; therapy was re-enabled.It was noted that the patient has another device within close proximity of the vns.Additional tablet data was received for review.From review of the decoder, burst watchdog timeouts were seen two additional times.The patient was able to have therapy enabled again and diagnostics were within normal limits after the events.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18124509
MDR Text Key328631616
Report Number1644487-2023-01629
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106
Was Device Available for Evaluation? No
Event Location Other
Initial Date Manufacturer Received 10/19/2023
Initial Date FDA Received11/13/2023
Supplement Dates Manufacturer Received04/05/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age20 YR
Patient SexMale
-
-