• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS TRANSFEMORAL BALLOON CATHETER; BALLOON AORTIC VALVULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS TRANSFEMORAL BALLOON CATHETER; BALLOON AORTIC VALVULOPLASTY Back to Search Results
Model Number 9350BC23
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Rupture (2208)
Event Date 10/30/2023
Event Type  Death  
Manufacturer Narrative
Per the instructions for use (ifu), cardiovascular injuries such as perforation or dissection of vessels, ventricle, myocardial or valvular structures, and hematoma are known potential adverse events associated with the overall thv procedure and may require intervention.According to the thv training manuals, risk factors for aortic dissection, hematoma, or annular rupture during the thv procedure include significant thv over sizing, severely obliterated sinuses of valsalva, porcelain aorta and/or presence of bulky calcification, and narrow calcified stj.In addition, advanced age, female gender, small body weight, and steroid dependency can also be contributing factors.The sapien valve relies on native valve calcium to securely anchor to the annulus.Despite this beneficial aspect of calcium, bulky calcium can increase the risk of calcific nodule displacement into the vasculature, which can lead to vascular injury.At times, the extent and distribution of calcium can impair ease of delivery of the valve, correct positioning of the valve, deployment of the valve, and procedural success.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate the cause of the event is unknown due to the limited information available.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.H3 other text: the device was not returned.
 
Event Description
As reported, it was a case of a 26mm sapien 3 valve, in aortic position by transfemoral approach.During procedure, a pre-dilatation was performed with a 23mm edwards balloon catheter and the patient's blood pressure drop.It was decided to implant the valve immediately.As an annular rupture was suspected, it was decided to implant an additional 26 mm valve was however, the patient expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS TRANSFEMORAL BALLOON CATHETER
Type of Device
BALLOON AORTIC VALVULOPLASTY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key18127452
MDR Text Key327989678
Report Number2015691-2023-17472
Device Sequence Number1
Product Code OZT
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K140241
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9350BC23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/30/2023
Initial Date FDA Received11/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age88 YR
Patient SexMale
-
-