Catalog Number 955403
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Event Description
It was reported the patient had excessive fluid removal of 500 ml with an ak 98 machine after dialysis treatment had completed.The target was 4kg; however, the actual was 4.5kg.It was not reported if an alarm had been triggered.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
E1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: h10: the device was not received for evaluation; however, the machine was evaluated on-site by a baxter qualified technician.Visual inspection of the machine was not reported.The technician confirmed the reported problem and identified the cause of the event as a defective uf measuring unit.The reported condition was verified.As the machine remained on-site, no further testing could be performed.Therefore, a more comprehensive device analysis could not be completed.To resolve the issue, the technician replaced the uf measuring unit.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 98 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18127770
MDR Text Key328421966
Report Number9616026-2023-00095
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414122966
UDI-Public(01)07332414122966
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/17/2023
Initial Date FDA Received11/13/2023
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received01/19/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-