• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.1
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2023
Event Type  Injury  
Manufacturer Narrative
A4 - unk.A5 - unk.A6 - unk.H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
 
Event Description
The reporter indicated that a 12.1mm, vticmo12.1, -13.0/1.5/094 (sphere/cylinder/axis), implantable collamer lens was implanted into the patient's right eye (od) on (b)(6) 2023.Lens rotation not associated to a low vault was observed, the lens was repositioned on (b)(6) 2023, but did not resolve the problem.On (b)(6) 2023 the lens was vertically replaced with a same length lens and the problem was resolved.Cause of the event is reported as unknown.
 
Manufacturer Narrative
Corrected data: b5 - it was reported that the lens was also repositioned on (b)(6) 2023.Claim#: (b)(4).
 
Manufacturer Narrative
H3 - device evaluation: the lens was returned dry in a microcentrifuge vial.Visual inspection found haptic bent.Dimensional inspection found the lens to be within specifications.Claim #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key18129708
MDR Text Key328024652
Report Number2023826-2023-04998
Device Sequence Number1
Product Code QBC
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberVTICMO12.1
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received11/14/2023
Supplement Dates Manufacturer Received01/24/2024
02/12/2024
Supplement Dates FDA Received02/09/2024
02/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL#SFC-45,LOT#UNK.; FOAMTIPPLUNGER MODEL#FTP,LOT#UNK.; INJECTOR MODEL#MSI-PF,LOT#UNK.
Patient Outcome(s) Required Intervention;
Patient Age26 YR
Patient SexFemale
-
-