Model Number DSX500S11
Device Problems Corroded (1131); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device has not been returned to the manufacture.
 
Event Description
The manufacturer became aware of allegations, that a dreamstation auto cpap states that the connection to the power inlet port is loose, also saw some rusty oxidation around the power inlet port.There was no report of serious patient harm or injury.No medical intervention was required by the patient.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation auto cpap unit.The allegation of the power connector of the unit was corroded and additionally findings rusty oxidation.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18131166
MDR Text Key328048630
Report Number2518422-2023-29549
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500S11
Device Catalogue NumberDSX500S11
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received11/14/2023
Supplement Dates Manufacturer Received10/16/2023
Supplement Dates FDA Received02/07/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-