MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL STEM 13DX145ST; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL STEM 13DX145ST; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Physical Resistance/Sticking (4012)
Patient Problem Insufficient Information (4580)
Event Date 10/04/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-03141 and 0001822565-2023-03142.G2: foreign: japan.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the patient underwent an initial surgery.Subsequently, the patient was revised approximately 31 years post implantation due to deformation of the acetabulum caused by prolonged use of these products.During the revision surgery, the original head and stem could not be pulled apart from each other.The surgery was planned to be performed with a 36 mm head, but a 36 mm ring-locked bipolar was used with cemented cup to complete the procedure.Attempts have been made and no further information has been provided.
 
Event Description
No further information at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d4; g3; h2; h3; h4; h6.Proposed component code: mechanical (g04)- stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported events.Medical records were not provided.A definitive root cause cannot be determined.No problem was found for the head and stem not disengaging, as per the cold weld memo they are designed to achieve long term fixation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 13DX145ST
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18134345
MDR Text Key328081633
Report Number0001822565-2023-03140
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K200823
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/1996
Device Model NumberN/A
Device Catalogue Number00652001300
Device Lot Number20170610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/18/2023
Initial Date FDA Received11/14/2023
Supplement Dates Manufacturer Received02/20/2024
Supplement Dates FDA Received02/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/24/1991
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-