• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 802018
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Manufacturer Narrative
The fsr replaced the flow module.Mechanical, electrical and visual testing was performed.The unit operated to the manufacturer's specifications.
 
Event Description
During field installation of the device, the field service representative (fsr) reported that the flow module failed testing.There was no patient involvement.
 
Manufacturer Narrative
During laboratory evaluation, the product surveillance technician (pst) connected the flow pod to lab use only (luo) testing equipment.There was no flow displayed on the screen when the pod was in use.The information on screen was as if there was no flow sensor attached.There were dashes displayed for the flow reading and a 'check sensor' message was displayed.With a luo generic printed circuit board (pcb) installed into the flow pod, the issue remained.With a luo application board installed into the flow pod, the flow displayed on the screen and the issue was corrected.The application pcb was the root cause of the reported complaint.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key18142875
MDR Text Key328983994
Report Number1828100-2023-00348
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000687
UDI-Public(01)00886799000687(11)230803
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number802018
Device Catalogue Number802018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received12/06/2023
Supplement Dates FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-