• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY S HBSAG REAGENT KIT; ANTIBODY TO HEPATITIS B SURFACE ANTIGEN (MOUSE MONOCLONAL IGG AND IGM)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY S HBSAG REAGENT KIT; ANTIBODY TO HEPATITIS B SURFACE ANTIGEN (MOUSE MONOCLONAL IGG AND IGM) Back to Search Results
Catalog Number 06P02-60
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2023
Event Type  Injury  
Event Description
The customer reported false reactive alinity s hbsag assay on cadaveric samples and provided the following information: reference = 1.0 s/co is reactive.On (b)(6) 2023, sample id (b)(6) initial result was 1.57 (reactive) and repeat results were 1.46 (reactive) & 1.26 (reactive).Nat result was negative.Tissue status: discarded patient information: 3-month-old male who was found unresponsive at home.Information regarding medications: 4 doses of epinephrine.Hepatitis b vaccination status was unknown, however the patient had 2 months of unspecified vaccines.Donor tissue discarded.
 
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Manufacturer Narrative
Section h3 was updated to indicate an evaluation is still in process.Section d4 - lot # was corrected h3 other text : an evaluation is still in process.
 
Event Description
The customer reported false reactive alinity s hbsag assay on cadaveric samples and provided the following information: reference = 1.0 s/co is reactive on (b)(6) 2023, sample id (b)(6) initial result was 1.57 (reactive) and repeat results were 1.46 (reactive) & 1.26 (reactive).Nat result was negative.Tissue status: discarded patient information: 3-month-old male who was found unresponsive at home.Information regarding medications: 4 doses of epinephrine.Hepatitis b vaccination status was unknown, however the patient had 2 months of unspecified vaccines.Donor tissue discarded.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review, and field data review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Device history review did not identify any potential non-conformances, non-conformances, or deviations with the complaint lot.Ticket search by lot did not identify an increase in complaint activity for the complaint issue.The ticket trending review did not identify any trends.The overall performance of the alinity s hbsag reagent was reviewed using field data from customers.A review of field data for the overall performance of lot 50489fn00 indicated the product was performing within product requirements.Labeling was reviewed and was found to address the customer reported issue.Based on the results of the investigation, alinity s hbsag reagent, lot number 50489fn00 met performance requirements and no systemic issue or product deficiency was identified.
 
Event Description
The customer reported false reactive alinity s hbsag assay on cadaveric samples and provided the following information: reference = 1.0 s/co is reactive.On (b)(6) 2023, sample id (b)(6) initial result was 1.57 (reactive) and repeat results were 1.46 (reactive) & 1.26 (reactive).Nat result was negative.Tissue status: discarded patient information: 3-month-old male who was found unresponsive at home.Information regarding medications: 4 doses of epinephrine.Hepatitis b vaccination status was unknown, however the patient had 2 months of unspecified vaccines.Donor tissue discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY S HBSAG REAGENT KIT
Type of Device
ANTIBODY TO HEPATITIS B SURFACE ANTIGEN (MOUSE MONOCLONAL IGG AND IGM)
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
finisklin business park
sligo F91VY 44
EI  F91VY44
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
finisklin business park
sligo F91VY 44
EI   F91VY44
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18142999
MDR Text Key328196776
Report Number3008344661-2023-00188
Device Sequence Number1
Product Code QHM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BL 125674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/07/2024
Device Catalogue Number06P02-60
Device Lot Number50489FN00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received11/28/2023
12/20/2023
Supplement Dates FDA Received12/14/2023
01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/03/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY S SYSTEM, 06P16-01, (B)(6); ALNTY S SYSTEM, 06P16-01, (B)(6); ALNTY S SYSTEM, 06P16-01,(B)(6)
Patient Outcome(s) Disability;
Patient Age3 MO
Patient SexMale
-
-