Model Number 7122Q/65
Device Problems Difficult to Fold, Unfold or Collapse (1254); Failure to Sense (1559); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2023
Event Type  Injury  
Event Description
During an in clinic follow up, a loss of sensing and increased threshold was observed on the right ventricular (rv) lead.Diagnostic imaging was performed and no anomalies were observed.A revision procedure was performed and when attempting to reposition the rv lead, the helix was unable to be extended.The rv lead was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported events were helix mechanism issue, loss of sensing and increase thresholds.As received, a complete lead was returned in one piece with helix found partially extended and clogged with dried blood/tissue.The reported event of helix mechanism issue was confirmed.X-ray inspection found overtorque of the inner coil at the connector region consistent with procedural damage.X-ray examination of the helix mechanism found no anomalies or distortion of the helix that would have contributed to helix mechanism issue reported in the field.After cleaning, the helix can be extended and retracted by applying torque directly at the connector pin.The full helix extension length was measured within specification.The cause of the reported event of helix mechanism issue was isolated to blood/tissue in the helix region and overtorqued of the inner coil.The reported events of loss of sensing and increase thresholds were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18143880
MDR Text Key328202281
Report Number2017865-2023-52550
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503341
UDI-Public05414734503341
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/65
Device Lot NumberS000090643
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/08/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received12/13/2023
Supplement Dates FDA Received12/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-